SERACCESS® - DISRUPTIVE TECHNOLOGY FOR THE MONITORING AND CONTROL OF CRITICAL BLOOD PARAMETERS

SERACCESS® MILESTONE 7A

Written by: Dagmar Brunner

Continuous automated blood sampling based on an animal model

Authors: Dagmar Brunner; Konstantinos Tzikas; Dr Reto Rapold; Dr med. vet. Karina Klein; Dr Beatrice Schär
Securecell AG/Vetsuisse-Fakultät ZH, MSRU/CABMM, May 2022

Keywords: Seraccess, automated blood sampling, animal model

 

Background

Automated blood glucose monitoring with Securecell’s SeraMaster proto-type was tested in a clinical setting as part of a feasibility study (see Milestone 6).

Several experiments on live sheep were defined to test automated blood sampling in vivo. In a first series of experiments, the automated sampling process, including dilution with NaCl, was compared with manual sampling.


Study Design

To evaluate the automated method, the bias (=mean deviation from the reference method) was calculated. For samples from the central vein the bias was 1.9 mg/dl (n=32) and for samples from the peripheral artery it was 1.7 mg/dl (n=40). All 72 data points (100%) are within the margin of error of ± 6.5 mg/dl. 
The following figure shows the regression analysis between SeraMaster glucose values (y-axis) and manual sampling (x-axis). The correlation coefficient (R2) is 0.9667, which means that the two methods are equivalent and therefore interchangeable.

 

Results

To evaluate the automated method, the bias (=mean deviation from the reference method) was calculated. For samples from the central vein the bias was 1.8 mg/dl (n=96) and for samples from the peripheral artery it was 0.8 mg/dl (n=104). All 200 data points (100%) are within the margin of error of ± 10.0 mg/dl. 
The following figure shows the regression analysis between SeraMaster glucose values (y-axis) and manual sampling (x-axis). The correlation coefficient (R2) is 0.9930, which means that the two methods are equivalent and therefore interchangeable.

REC_SER_Results animal test 7A
The regression analysis between SeraMaster glucose values (y-axis) and manual sampling (x-axis)

The following images illustrate the experimental setup in the stables of the university. At the end of the day, the laboratory animals were allowed to return to the pasture. 

experimental setup-MS7A The experimental setup in the stables of the university

 

Next steps

In the next series of experiments on sheep, automated blood sampling will be combined with automated glucose measurement.

Continuous automated blood sampling with the SeraMaster prototype is equivalent to the manual refer-ence method. The bias, a measure of the accuracy of the method, is 1.9 mg/dl (venous) and 1.7 mg/dl (arterial) compared to the reference method. The regression analysis shows that the two methods are interchangeable. Therefore, it makes sense to tackle the next series of experiments

© Copyright 2023 by Securecell AG. All rights reserved, including graphs and images. Lucullus and Numera are registered trademarks of Securecell AG, Switzerland.

Dagmar Brunner

Tags

Share